Set a second SwitchBox56 to start at 57 so you can run two SwitchBox56's for a total of 112 Electrodes
Plug in only 1 Switch Box 56 at a time to the SuperSting
Turn the SuperSting on and follow these steps from the Main Menu
- Test Menu (3)
- Program Address (5)
- Address = 57
- Serial Number = 1
- Press F3
- Press Period (.) to confirm
- Set these values next
- Test Menu (3)
- Program Address (5)
- Address=85
- Serial Number =2
- Press F3
- Press Period (.) to confirm
- Label with white tape and a marker that this Switch Box handles Electrodes 57 to 112
- To confirm that you have programmed 57 to 112, run the Relay Test from 57 to 112 as described here.
- Finally, label with white tape and a marker that the first Switch Box handles Electrodes 1 to 56 and connect both Switch Box 56's at the same time and run the Relay Test from 1 to 112
Related Links: How do I perform a SwitchBox Relay Test ?